AutoCalcimeter Model 442

Shale Tech Solutions SRL distribuidor exclusivo de FANN para la Argentina

The AutoCalcimeter determines the amount of calcium carbonate (CaCO3 or calcite) and calcium magnesium carbonate (CaMg(CO3)2 or dolomite) in alkaline earth carbonate samples – such as oil well cores or drill cuttings. Calcite or dolomite buildup in drilling fluids and water treatment process causes scaling problems, and data from the Fann AutoCalcimeter can help determine the most optimal chemical treatment.

This instrument is in compliance with ASTM D 4373-84: Standard Test Method for Calcium Carbonate Content in Soils.

Features

  • Automatic and accurate dispensing of HCl when test begins
  • Automatic stirring of sample with acid, by built-in stirring mechanism
  • Automatic recording of pressure readings and elapsed time internally and on USB flash drive
  • Automatic calcite and dolomite percentage reporting, with all calculations performed internally
  • Real-time clock stamps date and time
  • Data communication port for integration to computer

Application
Samples containing calcite and dolomite are reacted with 1 Normal HCl to measure the released CO2, in a sealed reaction cell. A calibration curve (determined using a pure calcium carbonate reagent) is used to relate the pressure to the weight of the calcium carbonate in the sample. Several weights of the sample are utilized, to ensure the most accurate curve, and pressures for the curve are measured and stored automatically when running the appropriate calibration sample size.

Recording and Interpreting Test Data
With a USB flash drive connected, the AutoCalcimeter will record the test data onto the USB drive in .csv file format. If the Flash drive is not present the data is still recorded in the internal memory. The AutoCalcimeter records the last 20 tests; this data can be downloaded using the transfer button on the keypad.

Test data can be interpreted using calibration curves for the pressure of calcite and dolomite. This material is provided in the included instruction manual.

Dimensions (W x D x H) 16 in x 9 in x 9.5in
Weight 15 lb
Power Supply 100 to 240 VAC, 50-60 hz, 75 watts
Ports Power, RS485 serial, USB
Data Communication USB

Ordering Information
Part No. 102452754 AutoCalcimeter Model 442
Includes plastic carrying case, mortar and pestle, pocket digital Weighing scale, CaCO3 compound, HCl, and instruction manual